Al2O3:C and LiF: Mg, Ti characterisations at 100–150 kV energy range for computed tomography dose measurement

Performance characterisation has been carried out of nanoDot™ OSLDs and TLD-100™, validating their use for diagnostic x-ray dose profile assessments. Investigations include optical annealing, signal depletion, signal fading, dose-response, sensitivity, and energy dependence. Both dosimeter types wer...

詳細記述

保存先:
書誌詳細
主要な著者: Omar, R.S., Hashim, S., Bradley, D.A., Karim, M.K.A., Kobayashi, I., Kadir, A.B.A., Hashim, A.
フォーマット: 論文
出版事項: Elsevier 2022
タグ: タグ追加
タグなし, このレコードへの初めてのタグを付けませんか!
id oai:psasir.upm.edu.my:100161
record_format eprints
spelling oai:psasir.upm.edu.my:100161 http://psasir.upm.edu.my/id/eprint/100161/ Al2O3:C and LiF: Mg, Ti characterisations at 100–150 kV energy range for computed tomography dose measurement Omar, R.S. Hashim, S. Bradley, D.A. Karim, M.K.A. Kobayashi, I. Kadir, A.B.A. Hashim, A. Performance characterisation has been carried out of nanoDot™ OSLDs and TLD-100™, validating their use for diagnostic x-ray dose profile assessments. Investigations include optical annealing, signal depletion, signal fading, dose-response, sensitivity, and energy dependence. Both dosimeter types were exposed using the Constant Potential Industrial X-ray (Model Philips MG 165) in Standard Radiation Qualities (RQT) procedure located in Nuclear Malaysia, Bangi. OSLD annealing using a 14 W compact fluorescent lamp showed an average signal loss of ∼93% as a result of 60 min of illumination. For dosimetric signal depletion, screened nanoDot™ OSLDs gave rise to the most favorable performance, with a mean signal loss of 1.0% per reading. In respect of signal fading, similar favorable performance was found for the screened nanoDot™ OSLDs, after a stabilisation period of 11–12 days post-irradiation, the average reading decreased by ∼1% over a further 17 days. For doses up to 500 mGy the TLD-100™ and screened nanoDot™ OSLDs both provide a highly linear response, with a regression coefficient of 0.999 in both cases. Linear energy dependence was found for RQT spectra from 100 to 150 kV. Elsevier 2022-10 Article PeerReviewed Omar, R.S. and Hashim, S. and Bradley, D.A. and Karim, M.K.A. and Kobayashi, I. and Kadir, A.B.A. and Hashim, A. (2022) Al2O3:C and LiF: Mg, Ti characterisations at 100–150 kV energy range for computed tomography dose measurement. Radiation Physics and Chemistry, 199. art. no. 110365. pp. 1-5. ISSN 0969-806X https://www.sciencedirect.com/science/article/pii/S0969806X2200408X?via%3Dihub 10.1016/j.radphyschem.2022.110365
institution UPM IR
collection UPM IR
description Performance characterisation has been carried out of nanoDot™ OSLDs and TLD-100™, validating their use for diagnostic x-ray dose profile assessments. Investigations include optical annealing, signal depletion, signal fading, dose-response, sensitivity, and energy dependence. Both dosimeter types were exposed using the Constant Potential Industrial X-ray (Model Philips MG 165) in Standard Radiation Qualities (RQT) procedure located in Nuclear Malaysia, Bangi. OSLD annealing using a 14 W compact fluorescent lamp showed an average signal loss of ∼93% as a result of 60 min of illumination. For dosimetric signal depletion, screened nanoDot™ OSLDs gave rise to the most favorable performance, with a mean signal loss of 1.0% per reading. In respect of signal fading, similar favorable performance was found for the screened nanoDot™ OSLDs, after a stabilisation period of 11–12 days post-irradiation, the average reading decreased by ∼1% over a further 17 days. For doses up to 500 mGy the TLD-100™ and screened nanoDot™ OSLDs both provide a highly linear response, with a regression coefficient of 0.999 in both cases. Linear energy dependence was found for RQT spectra from 100 to 150 kV.
format Article
author Omar, R.S.
Hashim, S.
Bradley, D.A.
Karim, M.K.A.
Kobayashi, I.
Kadir, A.B.A.
Hashim, A.
spellingShingle Omar, R.S.
Hashim, S.
Bradley, D.A.
Karim, M.K.A.
Kobayashi, I.
Kadir, A.B.A.
Hashim, A.
Al2O3:C and LiF: Mg, Ti characterisations at 100–150 kV energy range for computed tomography dose measurement
author_facet Omar, R.S.
Hashim, S.
Bradley, D.A.
Karim, M.K.A.
Kobayashi, I.
Kadir, A.B.A.
Hashim, A.
author_sort Omar, R.S.
title Al2O3:C and LiF: Mg, Ti characterisations at 100–150 kV energy range for computed tomography dose measurement
title_short Al2O3:C and LiF: Mg, Ti characterisations at 100–150 kV energy range for computed tomography dose measurement
title_full Al2O3:C and LiF: Mg, Ti characterisations at 100–150 kV energy range for computed tomography dose measurement
title_fullStr Al2O3:C and LiF: Mg, Ti characterisations at 100–150 kV energy range for computed tomography dose measurement
title_full_unstemmed Al2O3:C and LiF: Mg, Ti characterisations at 100–150 kV energy range for computed tomography dose measurement
title_sort al2o3:c and lif: mg, ti characterisations at 100–150 kv energy range for computed tomography dose measurement
publisher Elsevier
publishDate 2022
_version_ 1819301161680764928
score 13.4562235