Al2O3:C and LiF: Mg, Ti characterisations at 100–150 kV energy range for computed tomography dose measurement
Performance characterisation has been carried out of nanoDot™ OSLDs and TLD-100™, validating their use for diagnostic x-ray dose profile assessments. Investigations include optical annealing, signal depletion, signal fading, dose-response, sensitivity, and energy dependence. Both dosimeter types wer...